ORIGINAL ARTICLE
J Clin Imaging Sci 2018,  8:11

Evaluation of Orthopedic Metal Artifact Reduction Application in Three-Dimensional Computed Tomography Reconstruction of Spinal Instrumentation: A Single Saudi Center Experience


Department of Radiodiagnosis, Faculty of Medicine, Mansoura University, Egypt; Department of Medical Imaging, Sultan Bin Abdulaziz Humanitarian City, Riyadh 13571-6262, Kingdom of Saudi Arabia

Date of Submission09-Dec-2017
Date of Acceptance11-Feb-2018
Date of Web Publication12-Mar-2018

Correspondence Address:
Dr. Amir Monir Ali
Department of Radiodiagnosis, Faculty of Medicine, Mansoura University, Egypt. Department of Medical Imaging, Sultan Bin Abdulaziz Humanitarian City, Riyadh 13571-6262, Kingdom of Saudi Arabia

Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/2156-7514.227125

Rights and Permissions
   Abstract 


Aim of the Study: The aim of the study was to evaluate the commercially available orthopedic metal artifact reduction (OMAR) technique in postoperative three-dimensional computed tomography (3DCT) reconstruction studies after spinal instrumentation and to investigate its clinical application. Materials and Methods: One hundred and twenty (120) patients with spinal metallic implants were included in the study. All had 3DCT reconstruction examinations using the OMAR software after obtaining the informed consents and approval of the Institution Ethical Committee. The degree of the artifacts, the related muscular density, the clearness of intermuscular fat planes, and definition of the adjacent vertebrae were qualitatively evaluated. The diagnostic satisfaction and quality of the 3D reconstruction images were thoroughly assessed. Results: The majority (96.7%) of 3DCT reconstruction images performed were considered satisfactory to excellent for diagnosis. Only 3.3% of the reconstructed images had rendered unacceptable diagnostic quality. Conclusion: OMAR can effectively reduce metallic artifacts in patients with spinal instrumentation with highly diagnostic 3DCT reconstruction images.

Keywords: Orthopedic metal artifact reduction, spinal instrumentation, three-dimensional computed tomography


How to cite this article:
Ali AM. Evaluation of Orthopedic Metal Artifact Reduction Application in Three-Dimensional Computed Tomography Reconstruction of Spinal Instrumentation: A Single Saudi Center Experience. J Clin Imaging Sci 2018;8:11

How to cite this URL:
Ali AM. Evaluation of Orthopedic Metal Artifact Reduction Application in Three-Dimensional Computed Tomography Reconstruction of Spinal Instrumentation: A Single Saudi Center Experience. J Clin Imaging Sci [serial online] 2018 [cited 2018 Dec 11];8:11. Available from: http://www.clinicalimagingscience.org/text.asp?2018/8/1/11/227125




   Introduction Top


Spinal metallic implants are widely used in modern spinal surgery for increasing incidence and expanding etiologies. Magnetic resonance imaging evaluation is limited due to the magnetic severe susceptibility artifacts; so, computed tomography (CT) is now the method of choice for the postoperative assessment of spinal metallic instrumentation.[1],[2] However, the presence of spinal metal implants produces considerable artifacts on CT images, resulting in marked degradation of image quality and subsequent changes of CT attenuation values of the surrounding tissues, which seriously influences the reliability of the postoperative three-dimensional CT (3DCT) assessment of spinal instrumentation.[3] Hence, many studies investigated the metallic artifact reducing methods used in clinical practice including linear interpolation,[4] spline interpolation,[5] simultaneous algebraic reconstruction,[6] and simultaneous iterative reconstruction[7] techniques.

Orthopedic metal artifact reduction algorithm (OMAR) software is a commercial algorithm produced by Philips Health System, Cleveland, USA, that is used to remove metal artifacts by acquisition data alteration. Much iteration makes the new technique suitable for internal fixation of extended metal implants such as spinal plates, rods, and screws.[8] However, there are scarce reports regarding the evaluation of OMAR in postoperative 3DCT scans of spinal instrumentation.

The aim of this qualitative study was to assess the feasibility of the use of OMAR in the postoperative 3DCT reconstruction of spinal implants and investigate its clinical application.


   Materials and Methods Top


This study was conducted in Sultan Bin Abdulaziz Humanitarian City (SBAHC), Riyadh, Kingdom of Saudi Arabia. One hundred and twenty patients with spinal metallic instrumentation were included in this study from June to December 2017; 55 new cases (45.8%) and 65 cases coming for regular follow-up (54.2%). The segmental spinal regions were 9 cervical (7.5%), 101 dorsolumbar (84.2%), and 10 lumbar (8.3%). The study population included 108 (90%) males, with a median age of 31 and 12 (10%) females with the median age of 27. Etiology for spinal instrumentation included 49 (40.8%) posttraumatic fractures cases, 68 (56.7%) postscoliosis correction, and 3 (2.5%) post-tuberculosis (TB) cases. Informed consents and approvals of the SBAHC Scientific Research Ethical Committee were obtained for all patients before the CT examinations.

CT scans with 3D reconstruction were performed on a 64-slice CT scanner (Ingenuity; Philips Medical Systems, Cleveland, USA). Parameters used were helical CT protocol, iterative reconstruction: IDose Level 3 (range, 1–5), slice thickness 0.5 mm, slice interval 0.5 mm, matrix 512 × 512, collimation 64 × 0.625 mm, OMAR software, bone window (YD kernel, level 800 HU, width 2000 HU), and soft window (Soft kernel B). The degree of metallic artifacts and clarity of the scanned vertebrae of the 3D reconstruction images using OMAR were evaluated to determine their diagnostic acceptance.

Image quality evaluation was conducted on the source bone and soft-tissue windows as well as the 3D reconstruction images. A blinded consultant radiologist, with more than 15 years of experience in radiological diagnosis, conducted the subjective evaluations including severity of the spinal metallic artifacts, density of the related paraspinal muscles, clarity of the intermuscular fat planes, and definition of the scanned vertebral elements using a 5-point scale [Table 1].[1] Images scoring 3 or more were considered radiologically satisfactory.
Table 1: Radiological evaluation of orthopedic metal artifact reduction algorithm reconstructed image quality

Click here to view



   Results Top


In the produced spinal 3D-reconstructed images after OMAR application, the severity of the fixing metallic artifacts was significantly reduced, paraspinal muscles were more homogeneous in density, the intermuscular fat planes and planes between bone and instrumentation were clearer.

Images of 116 (96.7%), out of the 120 cases included in this study, were considered radiologically satisfactory to excellent. The score range of severity of instrumentation artifacts was 3–4 points, with the median of 4 points; and the range of definition of vertebrae was 3–5 points, with the median of 4 points. Thirty-six cases (30%) took the highest score of 5, 51 cases (42.5%) got Score 4, and 29 (24.2%) got Score 3. The four unacceptable cases (3.3%) were scored only one point for the ill-definition of the related vertebral bodies and posterior neural arches (2 cases) and inability to determine the intraspinal extension of TB-associated perispinal soft-tissue mass component (2 cases).


   Discussion Top


Artifacts caused by the presence of metallic implants in the CT scanned volume, such as spine, hip prostheses, or dental fillings, appear as dark and bright streaks across the reconstructed image.[9] Metallic artifacts can severely degrade the image quality and hence limit the diagnostic value of a CT scan.[10]

Metallic artifacts in CT scans are produced by beam hardening and photon starvation.[11] Photon starvation artifacts occur when the beam traverse materials with high-attenuation coefficients, which leads to a small amount of photons reaching the machine detectors and results in markedly noisy images.[12] The noise is multiplied in the reconstruction images, and the resulting streaks can be seen clearly. Beam hardening means that low energy photons are attenuated more than high energy photons when passing through the scanned object.[13] This effect is more prominent when the beam passes through very high-density materials such as metallic implants.[10]

Many solutions have been used to reduce metal artifacts; these are most often insufficient in terms of image quality for diagnosis. One approach is to increase the tube peak voltage; however, this may only reduce metallic artifact to a minor degree but may not improve imaging of larger implants. Increasing tube current, on the other hand, would lead to a higher radiation dose to the patient and may not have a considerable impact on image quality either.[11] Other solutions such as gantry tilting, or using lower attenuating materials in implants, may only be possible in specific situations and are effective to a minor extent. Multiple metallic artifact reduction algorithms working on raw projection data, such as modified iterative reconstruction and projection interpolation techniques, have been shown to be a more general and effective technique for reducing artifacts.[12],[13] Kidoh et al.[14] confirmed the value of OMAR with dental implants. Jeong et al.[15] discussed the value of the technique in improving the diagnostic accuracy of abdominal CT studies with metal artifacts. Li et al.[16] concluded that the algorithm could improve the quality of CT images taken for radiation therapy planning. Hu et al. conducted a study on a small scale of patients with the same conclusion.[1]

The OMAR algorithm used with Philips CT has previously been shown to improve CT number accuracy in images of orthopedic implants.[17] OMAR is an iterative projection modification method optimized for imaging orthopedic implants. The data acquired with the scanner are used as input into an iterative loop, where the output is a correction image that is subtracted from the input image. OMAR uses segmentation and replaces data points identified as metal with interpolated values. When using the OMAR technique, an uncorrected data volume is always automatically reconstructed.[12] In this study, we evaluated the use of this algorithm in postoperative 3D scans for spinal instrumentation and investigated its clinical application.

For elimination of metallic artifacts, the source data of the primary images, the sole implant images, and the scanned tissue images were compared and computed continuously until no metal data were detected, and the final images were reconstructed by filtered back projection technique.[18] Based on these calculations, metal artifacts are eliminated gradually; hence, OMAR is beneficial for sizable metallic implants. To increase the accuracy of metal artifact reduction, OMAR generates specific images for the metal based on CT attenuation on the original image, with a value of 0 for the metal density and 1 for the nonmetal densities; thus, the data acquisition is only for the metal domain with a value of 0. OMAR is supposed to have no side effects on the nonmetal parts of the images. After removal of the metal domain from the original acquired data, interpolation is applied to fill the lost data to reduce the variation in the produced images.[19]

The clinical application of OMAR for postspinal instrumentation was performed on 120 patient population coming for regular postoperative evaluations in SBAHC, one of the most prestigious rehabilitation centers in the Kingdom of Saudi Arabia, with two-international accreditations of CARF and JCI and one national accreditation of CBAHI. Cases included spinal instrumentation for posttraumatic [Figure 1] (49 cases), post-TB [Figure 2] (3 cases), postscoliosis correction [Figure 3] (68 cases) etiologies. [Figure 4] represents OMAR score 5, while [Figure 5], [Figure 6], [Figure 7], [Figure 8] represent lower OMAR scores. The results indicated that the images of 116 cases (96.7%) were scored 3 or more and were accepted by the radiologist, which confirmed the feasibility of OMAR in spinal instrumentation imaging. The degree of metallic artifacts and the definition of the adjacent vertebrae were all evaluated 4 points, impressive of little metallic artifacts with no obvious distortion. With the resulting clear 3D spinal reconstructions, the spinal surgeons could be sure about the relationship between vertebrae and the fixing hardware (e.g., the proper placing and depth of transpedicular screws, no injury of the related nerve roots, and no loosening or malplacement), evaluate the status of vertebrae surrounding the implant for osteoporosis or infection, and follow-up the healing process of fractures. Thus, this technique would help accurate postoperative assessment of spinal fixation of different reasons.[20] The 3D-reconstructed images were unsatisfactory in only four patients (3.3%) in this study; this could be attributed to the metal being too near to the surface [Figure 2] (one case) or too extensive (two cases of both anterior and posterior fixation and one case of posterior and lateral fixation), which can lead to faulty reconstruction by OMAR.
Figure 1: Case 1: A 42-year-old male posttraumatic vertebral fracture patient with complicated spinal instrumentation and extensive vertebral osteolysis. (a) Sagittal bone window reformat of the dorsolumbar spine using orthopedic metal artifact reduction algorithm showing the metallic implants with a clear definition of the vertebral details and minimal artifact streaks (Score 4). (b) Three dimensional computed tomography color-coded reconstruction with orthopedic metal artifact reduction, left posterior-lateral projection showing the fixing metallic rods with minimal intervening streak artifacts. (c) Three-dimensional computed tomography color-coded reconstruction with orthopedic metal artifact reduction, right posterior-lateral projection showing the extensive vertebral osteolysis with superior bone details and minimal streak artifacts.

Click here to view
Figure 2: Case 2: A 66-year-old male patient with spinal instrumentation for tuberculosis. (a) Sagittal soft-tissue window reformat of the dorsolumbar spine using orthopedic metal artifact reduction algorithm showing severe image degradation with obscuring artifact streaks (Score 1). (b) Axial bone window using orthopedic metal artifact reduction algorithm showing the metallic implants with marked veiling artifacts, ill-definition of the vertebral details and failure to delineate the extent of the perispinal tuberculosis lesion. (c) Three-dimensional computed tomography color-coded reconstruction with orthopedic metal artifact reduction, left posterior-lateral projection showing marked haziness of the bony details with marked streak artifacts.

Click here to view
Figure 3: Case 3: A 27-yar-old female patient with corrective spinal instrumentation for kyphoscoliotic deformity. (a) Sagittal bone window reformat of the dorsolumbar spine using orthopedic metal artifact reduction algorithm showing excellent vertebral details with almost absence of significant obscuring artifacts (Score 5). (b) Axial bone window using orthopedic metal artifact reduction algorithm, showing the metallic implants with clear vertebral and paraspinal muscle details. (c) Three-dimensional computed tomography color-coded reconstruction with orthopedic metal artifact reduction, left posterior-lateral projection showing the whole length of the corrective implant with clear anatomical relations.

Click here to view
Figure 4: Case 4: A 33-year-old patient with post-tuberculosis spinal fixation. (a) Sagittal bone window reformat of the dorsolumbar spine using orthopedic metal artifact reduction algorithm showing excellent vertebral details with almost absence of significant obscuring artifacts (Score 5). (b) Axial bone window using orthopedic metal artifact reduction algorithm, showing the tuberculous perispinal mass with clear intraspinal extension. (c) Three-dimensional computed tomography color-coded reconstruction with orthopedic metal artifact reduction, left posterior-lateral projection showing the fixing implant with well-defined tuberculosis bone destruction.

Click here to view
Figure 5: Case 5: A 36-year-old male patient with posttraumatic cervical spine fixation. (a) Sagittal bone window reformat of the cervical spine using orthopedic metal artifact reduction algorithm, showing internal fixation through a posterior approach with good vertebral details (Score 4). (b) Axial bone window using orthopedic metal artifact reduction algorithm, showing the good healing process of the vertebral fracture with no significant veiling artifacts. (c) Three-dimensional computed tomography color-coded reconstruction with orthopedic metal artifact reduction, right posterior-lateral projection showing the cervical instrumentation with satisfactory details.

Click here to view
Figure 6: Case 6: A 22-year-old male patient with post-L1 spinal fracture fixation. (a) Sagittal bone window reformat of the dorsolumbar spine using orthopedic metal artifact reduction algorithm, showing marked posterior cortical displacement of the fractured vertebra encroaching upon the spinal canal with satisfactory bone details (Score 3). (b) Axial bone window using orthopedic metal artifact reduction algorithm, showing satisfactory bone details yet, obscured adjacent paraspinal muscles. (c) Three-dimensional computed tomography color-coded reconstruction with orthopedic metal artifact reduction, left posterior-lateral projection showing the spinal canal encroachment with satisfactory details.

Click here to view
Figure 7: Case 7: A 55-year-old male patient with postfixation infection and hardware failure. (a) Sagittal bone window reformat of the dorsolumbar spine using orthopedic metal artifact reduction algorithm, showing marked loosening of the lower fixing pedicular screws with evident displacement and extensive osteolysis (Score 4). (b) Axial bone window using orthopedic metal artifact reduction algorithm showing the relation between the detached screws and the related vertebra. (c) Three-dimensional computed tomography color-coded reconstruction with orthopedic metal artifact reduction, left posterior-lateral projection showing the evident hardware failure and vertebral destruction.

Click here to view
Figure 8: Case 8: A 26-year-old female patient with nonunited vertebral fracture. (a) Sagittal bone window reformat of the cervical spine using orthopedic metal artifact reduction algorithm showing very clear details of the anterior spinal fixation (Score 5). (b) Axial bone window using orthopedic metal artifact reduction algorithm showing the gapping nonunited fractured posterior neural arch with the almost total absence of veiling artifacts. (c) Three-dimensional computed tomography color-coded reconstruction with orthopedic metal artifact reduction, anterior projection showing the good cervical alignment despite the nonunited fracture.

Click here to view


Philips recommends that this algorithm is not suitable for imaging of stents, external fixation metals, implanted devices near skin surface, metals near air pockets, and surgical screws or clips. There are some cases where OMAR should be avoided as outlined in the contraindications section.[21],[22] These commonly occur when the metal is in close proximity to air or lung tissue or small metal object (e.g. stents) within iodinated contrast. The radiologist should be encouraged to cross reference the uncorrected images with the OMAR dataset when there are unsatisfactory construction images. Since the system will always reconstruct both sets of images whenever OMAR is selected, the uncorrected images are readily available.[23]

To have a complete evaluation of this OMAR technique, it would be of interest to perform a quantitative image analysis of CT numbers as a complement to the qualitative analysis carried out in this study. Other visual grading studies of specific spinal regions, cervical, dorsal or lumbosacral images, or based on the etiology for spinal instrumentation, would be complementary to this study.


   Conclusion Top


From this postspinal instrumentation study on 120 patients, it was concluded that, using the OMAR algorithm, combined with 3D reconstruction, much improved the produced images in terms of image quality, diagnostic accuracy, and better anatomical structure visualization, avoiding subjective density overrides of artifact regions on uncorrected images. Further quantitative image analysis studies for OMAR evaluation would be of interest.

Declaration of patient consent

The authors certify that they have obtained all appropriate patient consent forms. In the form the patient(s) has/have given his/her/their consent for his/her/their images and other clinical information to be reported in the journal. The patients understand that their names and initials will not be published and due efforts will be made to conceal their identity, but anonymity cannot be guaranteed.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
   References Top

1.
Hu Y, Pan S, Zhao X, Guo W, He M, Guo Q,et al. Value and clinical application of orthopedic metal artifact reduction algorithm in CT scans after orthopedic metal implantation. Korean J Radiol 2017;18:526-35.  Back to cited text no. 1
    
2.
Buckwalter KA, Rydberg J, Kopecky KK, Crow K, Yang EL. Musculoskeletal imaging with multislice CT. AJR Am J Roentgenol 2001;176:979-86.  Back to cited text no. 2
[PUBMED]    
3.
White LM, Buckwalter KA. Technical considerations: CT and MR imaging in the postoperative orthopedic patient. Semin Musculoskelet Radiol 2002;6:5-17.  Back to cited text no. 3
[PUBMED]    
4.
Kim M, Choi YS, Kim H, Choi NH. Postoperative evaluation after anterior cruciate ligament reconstruction: Measurements and abnormalities on radiographic and CT imaging. Korean J Radiol 2016;17:919-30.  Back to cited text no. 4
[PUBMED]    
5.
Rinkel J, Dillon WP, Funk T, Gould R, Prevrhal S. Computed tomographic metal artifact reduction for the detection and quantitation of small features near large metallic implants: A comparison of published methods. J Comput Assist Tomogr 2008;32:621-9.  Back to cited text no. 5
[PUBMED]    
6.
Abdoli M, Ay MR, Ahmadian A, Zaidi H. A virtual sinogram method to reduce dental metallic implant artefacts in computed tomography-based attenuation correction for PET. Nucl Med Commun 2010;31:22-31.  Back to cited text no. 6
[PUBMED]    
7.
Andersen AH, Kak AC. Simultaneous algebraic reconstruction technique (SART): A superior implementation of the art algorithm. Ultrason Imaging 1984;6:81-94.  Back to cited text no. 7
[PUBMED]    
8.
Gilbert P. Iterative methods for the three-dimensional reconstruction of an object from projections. J Theor Biol 1972;36:105-17.  Back to cited text no. 8
[PUBMED]    
9.
Metal Artifact Reduction for Orthopedic Implants (O-MAR). Available from: http://www.clinical.netforum.healthcare.philips.com/us_en/Explore/White-Papers/CT/Metal-Artifact-Reduction- for-Orthopedic-Implants- (O-MAR)#. [Last accessed on 2017 Nov 22].  Back to cited text no. 9
    
10.
Barrett JF, Keat N. Artifacts in CT: Recognition and avoidance. Radiographics 2004;24:1679-91.  Back to cited text no. 10
[PUBMED]    
11.
Lee MJ, Kim S, Lee SA, Song HT, Huh YM, Kim DH,et al. Overcoming artifacts from metallic orthopedic implants at high-field-strength MR imaging and multi-detector CT. Radiographics 2007;27:791-803.  Back to cited text no. 11
[PUBMED]    
12.
Nakae Y, Sakamoto K, Minamoto T, Kamakura T, Ogata Y, Matsumoto M,et al. Clinical evaluation of a newly developed method for avoiding artifacts caused by dental fillings on X-ray CT. Radiol Phys Technol 2008;1:115-22.  Back to cited text no. 12
[PUBMED]    
13.
Yazdi M, Gingras L, Beaulieu L. An adaptive approach to metal artifact reduction in helical computed tomography for radiation therapy treatment planning: Experimental and clinical studies. Int J Radiat Oncol Biol Phys 2005;62:1224-31.  Back to cited text no. 13
[PUBMED]    
14.
Kidoh M, Nakaura T, Nakamura S, Tokuyasu S, Osakabe H, Harada K,et al. Reduction of dental metallic artefacts in CT: Value of a newly developed algorithm for metal artefact reduction (O-MAR). Clin Radiol 2014;69:e11-6.  Back to cited text no. 14
[PUBMED]    
15.
Jeong S, Kim SH, Hwang EJ, Shin CI, Han JK, Choi BI,et al. Usefulness of a metal artifact reduction algorithm for orthopedic implants in abdominal CT: Phantom and clinical study results. AJR Am J Roentgenol 2015;204:307-17.  Back to cited text no. 15
    
16.
Li H, Noel C, Chen H, Harold Li H, Low D, Moore K,et al. Clinical evaluation of a commercial orthopedic metal artifact reduction tool for CT simulations in radiation therapy. Med Phys 2012;39:7507-17.  Back to cited text no. 16
[PUBMED]    
17.
Hilgers G, Nuver T, Minken A. The CT number accuracy of a novel commercial metal artifact reduction algorithm for large orthopedic implants. J Appl Clin Med Phys 2014;15:4597.  Back to cited text no. 17
[PUBMED]    
18.
Andersson KM, Nowik P, Persliden J, Thunberg P, Norrman E. Metal artefact reduction in CT imaging of hip prostheses – An evaluation of commercial techniques provided by four vendors. Br J Radiol 2015;88:20140473.  Back to cited text no. 18
    
19.
Pessis E, Campagna R, Sverzut JM, Bach F, Rodallec M, Guerini H,et al. Virtual monochromatic spectral imaging with fast kilovoltage switching: Reduction of metal artifacts at CT. Radiographics 2013;33:573-83.  Back to cited text no. 19
[PUBMED]    
20.
Brook OR, Gourtsoyianni S, Brook A, Mahadevan A, Wilcox C, Raptopoulos V,et al. Spectral CT with metal artifacts reduction software for improvement of tumor visibility in the vicinity of gold fiducial markers. Radiology 2012;263:696-705.  Back to cited text no. 20
    
21.
Meyer E, Raupach R, Lell M, Schmidt B, Kachelriess M. Normalized metal artifact reduction (NMAR) in computed tomography. Med Phys 2010;37:5482-93.  Back to cited text no. 21
[PUBMED]    
22.
Wang G, Frei T, Vannier MW. Fast iterative algorithm for metal artifact reduction in X-ray CT. Acad Radiol 2000;7:607-14.  Back to cited text no. 22
[PUBMED]    
23.
Huang JY, Kerns JR, Nute JL, Liu X, Balter PA, Stingo FC,et al. An evaluation of three commercially available metal artifact reduction methods for CT imaging. Phys Med Biol 2015;60:1047-67.  Back to cited text no. 23
[PUBMED]    

 
   Authors Top

Dr. Amir Monir Ali


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6], [Figure 7], [Figure 8]
 
 
    Tables

  [Table 1]



 

Top
Previous article   Next article
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Conclusion
    References
    Authors
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed824    
    Printed41    
    Emailed0    
    PDF Downloaded165    
    Comments [Add]    

Recommend this journal