ORIGINAL ARTICLE
Year : 2018  |  Volume : 8  |  Issue : 1  |  Page : 9

Radiation reduction in low dose pulsed fluoroscopy versus standard dose continuous fluoroscopy during fluoroscopically-guided lumbar punctures: A prospective controlled study


Department of Radiology, Penn State Health Milton S. Hershey Medical Center, Hershey, PA, USA

Correspondence Address:
Dr. Shyam Sabat
Department of Radiology, H066, Penn State Milton S Hershey Medical Center, Hershey, PA 17033
USA
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/2156-7514.227044

Rights and Permissions

Purpose: The purpose of this study was to evaluate radiation dose reduction in fluoroscopically guided lumbar punctures (FGLP) using “pulsed fluoroscopy in a low dose mode” compared with the commonly used “continuous fluoroscopy in a standard dose mode” while maintaining the technical success. Materials and Methods: Thirty-five consecutive patients who underwent FGLP divided randomly to seventeen patients in the control group with standard dose continuous FGLP and eighteen patients in the study group with low-dose low-frame-rate of 3 frames per second (fps) FGLP. Entrance surface dose measurements from a dosimeter device were recorded as well as the dose area product (DAP). Results: A total of 35 patients with average age of 52 years (range: 15–87 [±17 standard deviation [SD]) were evaluated. Average entrance surface dose of the study group was significantly lower (3.81 mGy [range: 0.21–11.14, [±2.8 SD]) compared with the control group (22.45 mGy [range: 1.23–73.44, [±19.41 [SD]). The average DAP of the study group (10 mGy·cm2 [range: 1–41, [±9.8 SD]) was also significantly lower than the control group (65 mGy·cm2 [range: 5–199, [±53 SD]). Success rate was similar between the study and control groups. Conclusion: Low dose pulsed fluoroscopy of 3 fps significantly reduces radiation exposure by about 600% compared with standard dose continuous fluoroscopy in FGLP. Utilizing this radiation saving strategy will allow to dramatically reduce radiation exposure, without impacting the technical success rate.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed751    
    Printed23    
    Emailed0    
    PDF Downloaded143    
    Comments [Add]    

Recommend this journal