Un-re-sheath-able Misaligned pCONus Device: Case Report of a Unique Complication

Krishna Prasad Bellam PremnathGigy KuruttukulamAnil Kumar Shivan
Department of Neurointerventions, Rajagiri Hospital, Kochi, Kerala, India
Date of Submission: 04-May-2017, Date of Acceptance: 04-Jul-2017, Date of Web Publication: 27-Jul-2017.
Corresponding Author:
Corresponding Author

Krishna Prasad Bellam Premnath

C4, First Block, AVR Apartments, Thammannan Road, Salem - 636 009, Tamil Nadu, India.
E-mail: krishnaprasadir@gmail.com

Corresponding Author:
Corresponding Author

Krishna Prasad Bellam Premnath

C4, First Block, AVR Apartments, Thammannan Road, Salem - 636 009, Tamil Nadu, India.
E-mail: krishnaprasadir@gmail.com

DOI: 10.4103/jcis.JCIS_36_17 Facebook Twitter Google Linkedin

ABSTRACT


pCONus is a stent-like endovascular device which aids in retention of coils within wide-necked bifurcation aneurysms. It is retrievable even after complete deployment and is detached electrolytically. The pCONus aided coiling of wide-necked bifurcation aneurysms has a high technical success rate and a good safety profile. Different complications have been described in literature with the usage of pCONus. This case report describes a unreported complication of inappropriately deployed pCONus device.
Keywords: Aneurysm, Endovascular Coiling, PCONus Device, Waffle-cone Technique

This is an open-access article distributed under the terms of the Creative Commons Attribution-NonCommercial-Share Alike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms.